Introduction protocol


Based on Hänsch, Sebastian 27.11.2018

Start

User sample

What are the samples? What are you interested in testing? What functions do we need from this microscope?

Safety

  • Genetic safety - ask for "Excerpt/wrapper of Formblatt Z" for each new project - to describe GMOs, send to CAi
  • Chemical safety - corrosion, toxic substances
  • Laser safety - no mirrors, opening boxes, etc. - and complete the laser safety course!
  • Log-book writing and error reporting, how to do.

Booking

  • Booking policies (show website: 3 h slots, 4-wk in advance max)
  • Don't book and then just before, remove a massive session
  • We use the booking system to generate bills

Incubator

  • We keep it at 21.5°C, all the time (do not leave it open!)
  • Other cooling: right-side of the table is at 17°C, all the time, for the objective turret
  • Leave coolings on!

Switch-on

  • Buttons in numbered sequence
  • Pause between buttons, do you need the fluorescent lamp?

Log-in

  • TCS user - machine - DMI8 stage
  • Generally non-active (but optionally on):
  • Resonant scanner (greatly increase scan speed), STED (high resolution), AFC (adaptive focus control - good for keeping time course in focus)
  • Initialize table (for multiple locations imaging; raise microscope arm first, objectives go down automatically)

Software - lasers

  • Laser config: STED vs. non-STED
  • Which lasers do we have? What do they do? WLL (pulsed, pick wavelength, gating), Argon (more powerful, fan cooling)
  • How to switch them off (especially Argon)
  • The fluorescent lamp needs at least 30 minutes between off-on

Touch panel

  • Also has a rotary knob with forward = toward sample
  • Objective switching, fluorecent light / cubes on-off, pick, transmitted light
  • Immersion of objectives: which one to choose?
  • Save positions using touch pad: working/safe positions. Careful not to crash automatically...

Working distance

The 100x objective, NA=1.40, works very close to the coverslip, so pay special attention

Table inlet

  • Is for fine z-adjustments, eg. active when xz-scanning
  • Do not lean on, or push down hard on!

Sample placement/preparation

  • Table sample holder: move holders to accomodate sample - leave open to allow raising sample
  • Objective selection + apply immersion oil
  • Focus finding: see droplet become wider, use fluorescence/DIC

Software - overview

  • Left: acquisition settings, pixel size, speed etc.
  • Middle: beam path, filter settings, detectors etc.
  • Right: images already taken, current scan displayed

Beam path

  • Go through beam path on-screen, what it schematically represents
  • Lasers: switch on check boxes, switch between classic/spectrum-picker views. Start with 2-5% power
  • Notch filters: picking laser lines at notch filter wavelengths is preferable over exactly matching the dye's excitation spectrum.
  • Detectors: PMT (robust, noisy) vs. HyD (sensitive, advanced functions: gating, photon counting, low noise)

Laser/filter settings

  • Picking emission spectrum: Keep laser out of range, at least 10 nm
  • Transmitted light on: in parallel to laser scanning -can just activate in any sequential scan
  • Show dye spectrum and "dye assistant" (help with setting up scans)

Acquisition

  • Rotary knobs on touch-pad, configurable too!
  • Autofocus can find focus automatically, but isn't used often (by "hand", rather)
  • Live mode (quick and dirty), doesn't take into account detailed settings

Live

  • We see nothing at first - why? PMT used, needs gain (600-800), adjust contrast in histogram to keep up
  • Motor correction collar: What does it do and why?
  • Motor correction can set up on control panel, and can adjust until the signal is brightest (eg. use XZ scanning)
  • Use "range indicator" to view live, and to optimize motorized correction - in steps, judge when signal is going down again

PMT vs HyD

Demonstrate the difference between using PMTs and HyDs:

  • Grab the schematic display (with photon cascades etc.) that Leica made
  • Point out low HyD background noise (range indicator)
  • Point out gating option to further reduce eg. reflections (use coverslip laser reflection pattern)
  • Mention the capacity for overload (HyD starts to beep if so), and contact CAi
  • HyDs employ a stretching factor when changing gain: intensity values are multiplied and displayed

  • Can simulate a low-SNR situation by just using low laser, and rescuing the situation with HyD settings

Fine tuning acquisition

  • Pixel size is important - talk about the extremes, and what's wrong with those
  • Optimally, pixel size should be 3-4 times smaller than the expected resolution (which is roughly 1/2 the emission wavelength)
  • Pinhole: can change values to increase fluorescent signal intensity, at the cost of resolution
  • Averaging is employed (not in live) to counteract noise, such as PMT detector noise
  • Accumulation is employed (also in live) in situations with low signal intensity, but also low noise

HyD usage options

  • "photon counting mode" + accumulation (not averaging)
  • "standard mode" + averaging
  • "gating" (default: >300 ps) to counter reflections, autofluorescence
  • Optimise the signal for each channel using laser intensity, detector gain, scanning speed, averaging/accumulation, zoom, and display range
  • Optional: smooth rendering (leica display option) can be disabled, r/clicking the channel image

Multi-channel experiments

  • Activate T-PMT → immediately have a multi-channel experiment
  • Mention sequential scanning (line, frame)
  • Add channel ("Seq")
  • Cannot change detector order, physically arranged that way
  • Add a laser line + detector window, point out you need to switch them on/off in each sequential scan
  • Dye Assistant does these things automatically, but needs scrutiny
  • Can use Dye Assistant to point out potential cross-talk and bleedthrough issues (spectra are visualised)
  • The real way to address bleedthrough/crosstalk problems: USE CONTROL SAMPLES! With eg. only one colour of fluorophore expressed.
  • Then, suspicious signal can be evaluated; does it occur in the sample itself? or because of the other fluorophore? at all?

Re-using settings

Acquisition settings (laser, filter settings, speed etc.) can be set up from scratch or re-used:

  • from an existing image
  • from saved user settings on the microscope computer
  • sequential scan settings (eg. multi-channel combinations) need to be saved with a separate button

Z-stacks

  • Use gentle imaging conditions, and the control-panel Z-controls, define top and bottom of the stack
  • The knobs on the microscope control the height of the whole table, the knob on the control panel controls the smaller Z-inlet
  • Change step-size, best is system optimized, but can reduce them a bit

Time series/focus control

  • Define total experiment time, and intervals between acquisitions
  • The series will use the currently active acquisition settings (speed, format, etc.)
  • Now the Automated Focus Control (AFC) becomes relevant, which uses IR laser reflection to hold-current-position over time
  • "Show autofocus channel" can show what's going on there

HyVolution

  • PSF always messes up your image, but can be partially countered using deconvolution
  • HyVolution slider drags between best speed and best resolution, for deconvolution
  • Enter which mounting medium was used
  • Change from aggressive-standard-weak setting; test same acquisition region for the differences
  • Try to overdo it, get artifacts, to show what it looks like when going wrong

Saving

  • Can save data locally during the experiment,
  • Then transfer to the web drive (Daten) at the end of the session
  • We do not take responsibility for data locally saved!
  • Can make "collections" out of images to create separate LIF files
  • Keep LIF files, use FIJI to analyze them, export things from there

Switch-off

  • Depends on next user: if same-day, can keep system on
  • Clean the objectives with a few wipes using a Kimtech-wipe
  • Lasers off in software, wait for Argon laser to cool (can hear)
  • Then, switch off PC, and reverse-order the power buttons on the panel
  • Close the incubator and write session details in log-book






  • Keine Stichwörter