Versionen im Vergleich

Schlüssel

  • Diese Zeile wurde hinzugefügt.
  • Diese Zeile wurde entfernt.
  • Formatierung wurde geändert.

Introduction protocol

...

Based on Hänsch, Sebastian 27.11.2018

Start

Tipp
titleUser sample

What are the samples? What are you interested in testing? What functions do we need from this microscope?

Safety

  • Genetic (safety - ask for "Excerpt/wrapper of Form Formblatt Z" for each new project - to describe GMOs, send to CAi
  • Chemical ( safety - corrosion, toxic )substances
  • Laser ( safety - no mirrors, opening boxes, etc)

Booking

  • . - and complete the laser safety course!
  • Log-book writing (can refer to formblatt Z excerpt/number)and error reporting, how to do.

Booking

  • Booking policies (show website: 3 h slots, 4-wk in advance max)

...

titleSample
  • Don't book and then just before, remove a massive session
  • We use the booking system to generate bills

...

Incubator

  • we keep it at 21.5C5°C, all the time (do not leave it open!)
  • other cooling: right-side of the table = 17C, 24/7 is at 17°C, all the time, for the objective turret
  • leave coolings on!

Switch-on

  • Buttons in numbered sequence
  • Pause between buttons, do you need the fluorescent lamp?

...

  • TCS user - machine - DMI8 stage
  • Generally non-active (but optionally on):
  • Resonant scanner (greatly increase scan speed), STED (high resolution), AFC (adaptive focus control - good for keeping time course in focus)
  • Initialize table (for multiple locations imaging; raise microscope arm first, objectives go down automatically)

...

titleWorking distance

...

Software - lasers

  • Laser config: STED vs. non-STED
  • Which lasers do we have? What do they do? WLL (pulsed, pick wavelength, gating), Argon (more powerful, fan cooling)
  • How to switch them off (especially Argon)
  • The fluorescent lamp needs at least 30 minutes between off-on

Touch panel

  • Also has a rotary knob with forward = toward sample
  • Objective switching, fluorecent light / cubes on-off, pick, transmitted light
  • Immersion of objectives: which one to choose?
  • Save positions using touch pad: working/safe positions. Careful not to crash automatically...

Tipp
titleRotary knobsWorking distance

The 100x objective, NA=1.40, works very close to the coverslip, so pay special attentionRotaryknobs on the touch pad and microscope - rotate away from you = move objective closer to sample

Table inlet

  • Is for fine z-adjustments, eg. active when xz-scanning
  • Do not lean on, or push down hard on!

...

  • Table sample holder: move holders to accomodate sample - leave open to allow raising sample
  • Objective selection + apply immersion oil
  • Focus finding: see droplet become wider, use fluorescence/DIC

Software - overview

  • Left: acquisition settings, pixel size, speed etc.
  • Middle: beam path, filter settings, detectors etc.
  • Right: images already taken, current scan displayed

Beam path

  • Go through beam path on-screen, what it schematically represents
  • Lasers: switch on check boxes, switch between classic/spectrum-picker views. Start with 2-5% power
  • Notch filters: picking laser lines at notch filter wavelengths is preferable over exactly matching the dye's excitation spectrum.
  • Detectors: PMT (robust, noisy) vs. HyD (sensitive, advanced functions: gating, photon counting, low noise)

Laser/filter settings

  • Picking emission spectrum: Keep laser out of range, at least 10 nm
  • Transmitted light on: in parallel to laser scanning -can just activate in any sequential scan
  • Show dye spectrum and "dye assistant" (help with setting up scans)

Acquisition

  • Rotary knobs on touch-pad, configurable too!
  • Autofocus can find focus automatically, but isn't used often (by "hand", rather)
  • Live mode (quick and dirty), doesn't take into account detailed settings

Live

  • We see nothing at first - why? PMT used, needs gain (600-800), adjust contrast in histogram to keep up
  • Motor correction collar: What does it do and why?
  • Motor correction can set up on control panel, and can adjust until the signal is brightest (eg. use XZ scanning)
  • Use "range indicator" to view live, and to optimize motorized correction - in steps, judge when signal is going down again
Tipp
titlePMT vs HyD

Demonstrate the difference between using PMTs and HyDs:

  • Grab the schematic display (with photon cascades etc.) that Leica made
  • Point out low HyD background noise (range indicator)
  • Point out gating option to further reduce eg. reflections (use coverslip laser reflection pattern)
  • Mention the capacity for overload (HyD starts to beep if so), and contact CAi
  • HyDs employ a stretching factor when changing gain: intensity values are multiplied and displayed

  • Can simulate a low-SNR situation by just using low laser, and rescuing the situation with HyD settings

Fine tuning acquisition

  • Pixel size is important - talk about the extremes, and what's wrong with those
  • Optimally, pixel size should be 3-4 times smaller than the expected resolution (which is roughly 1/2 the emission wavelength)
  • Pinhole: can change values to increase fluorescent signal intensity, at the cost of resolution
  • Averaging is employed (not in live) to counteract noise, such as PMT detector noise
  • Accumulation is employed (also in live) in situations with low signal intensity, but also low noise
Tipp
titleHyD usage options
  • "photon counting mode" + accumulation (not averaging)
  • "standard mode" + averaging
  • "gating" (default: >300 ps) to counter reflections, autofluorescence
  • Optimise the signal for each channel using laser intensity, detector gain, scanning speed, averaging/accumulation, zoom, and display range
  • Optional: smooth rendering (leica display option) can be disabled, r/clicking the channel image

Multi-channel experiments

  • Activate T-PMT → immediately have a multi-channel experiment
  • Mention sequential scanning (line, frame)
  • Add channel ("Seq")
  • Cannot change detector order, physically arranged that way
  • Add a laser line + detector window, point out you need to switch them on/off in each sequential scan
  • Dye Assistant does these things automatically, but needs scrutiny
  • Can use Dye Assistant to point out potential cross-talk and bleedthrough issues (spectra are visualised)
  • The real way to address bleedthrough/crosstalk problems: USE CONTROL SAMPLES! With eg. only one colour of fluorophore expressed.
  • Then, suspicious signal can be evaluated; does it occur in the sample itself? or because of the other fluorophore? at all?
Tipp
titleRe-using settings

Acquisition settings (laser, filter settings, speed etc.) can be set up from scratch or re-used:

  • from an existing image
  • from saved user settings on the microscope computer
  • sequential scan settings (eg. multi-channel combinations) need to be saved with a separate button

Z-stacks

  • Use gentle imaging conditions, and the control-panel Z-controls, define top and bottom of the stack
  • The knobs on the microscope control the height of the whole table, the knob on the control panel controls the smaller Z-inlet
  • Change step-size, best is system optimized, but can reduce them a bit

Time series

  • Define total experiment time, intervals between acquisitions
  • Now the AFC becomes relevant, which uses IR laser reflection to hold-current-position over time
  • "show autofocus channel" can show what's going on there

HyVolution

  • PSF always messes up your image, but can be partially countered using deconvolution
  • HyVolution slider drags between best speed and best resolution, for deconvolution
  • Enter which mounting medium was used
  • Change from aggressive-standard-weak setting; test same acquisition region for the differences
  • Try to overdo it, get artifacts, to show what it looks like when going wrong

Saving

  • Can save data locally during the experiment,
  • Then transfer to the web drive (Daten) at the end of the session
  • We do not take responsibility for data locally saved!
  • Can make "collections" out of images to create separate LIF files
  • Keep LIF files, use FIJI to analyze them, export things from there

Switch-off

  • Depends on next user: if same-day, can keep system on
  • Clean the objectives with a few wipes using a Kimtech-wipe
  • Lasers off in software, wait for Argon laser to cool (can hear)
  • Then, switch off PC, and reverse-order the power buttons on the panel
  • Close the incubator and write session details in log-book